분산라만분광계 (Raman)
사진 |
![]() |
---|---|
영문명/단축명 | Raman Microscope / Raman |
모델명 |
LabRam Soleil |
설치장소 | 유기분석실 / [81601]공동기기원실험실 |
제작사 | Horiba |
도입년도/가격 | 2024. 2월 / 299,200,000원 |
담당자 연락처/e-mail | 031-299-6743 / 81601@skku.edu |
♦ Features
A high-performance confocal Raman system offering multimodal imaging (bright/dark field, PL, TERS, NanoPL) in a compact <1 m² design.
Features include:
• QScan™ for 3D confocal imaging
• SWIFT™ & SmartSampling™ for ultrafast mapping
• Auto-alignment & EasyImage™ for full automation
• Up to 4 lasers, 6 filters, and 30 cm⁻¹ low-frequency detection
• LabSpec 6 software with compliance (FDA/GMP), multivariate analysis, and particle ID
♦ Specification
Item | Specification | Description |
---|---|---|
Wavelength range | UV-VIS-NIR | Broadband high throughput achromatic mirror-based system, optimized from 300 nm to 1600 nm without changing optics. |
Standard Laser Wavelengths | 325, 405, 473, 532, 638, 785 nm | Typical lasers for Raman and PL. Other wavelengths upon request. |
Spectrometer scanning speed | Up to 400 nm/s | With 600g/mm grating, mounted on standard Turbodrive 4-grating turret, for fast Raman, PL spectral map acquisitions. |
Number of gratings | Unlimited | 4 grating exchangeable motorized turret. |
Fast Imaging | < 1 ms/spectrum | SWIFT, SWIFT XS EMCCD, SWIFT Repetitive, SWIFT Extended Range and SmartSampling for ultrafast imaging. |
3D confocal slicing with laser lightsheet illumination | QScan (patent) | Typical 2 μm thick 100 x 100 μm² laser lightsheets for large area 2D/3D confocal imaging. |
Standard wavenumber cut-off | 30 cm-1 | With edge filters for 532, 638 and 785 nm wavelengths, injection rejection, >99% transmission. |
Laser auto-alignment procedure | 15 s | Optional ultrafast sample-independent laser auto-alignment allowing remote maintenance. |
Lasers | Up to 4 built-in solid-state lasers + 1 external laser | NUV to NIR wavelengths available. |
Spectroscopic motorized modalities | Up to 6 laser filters | Raman, PL, Super Low Frequency… |
Rayleigh Filter orientation | Individual filter computer controlled | Factory preset & user adjustable angle to adapt with sample reflectivity. |
AFM / SEM coupling | Yes | Built-in horizontal exit for direct Atomic Force microscope coupling, optional fiber entrance for SEM coupling with RCLUE. |
Operating temperature range | 18–28 °C | Non condensing |
Dimensions (W x D x H in mm) Weight (kg) |
< 900 x 800 x 810 120 kg |
~1 cubic meter, including lasers, CDRH enclosure, electronics and cooling to save space in your laboratory |
Power consumption | < 600 W | Eco-friendly and safe design, with built-in class 1 enclosure, low power consumption electronics and cooling, 1 power cable EU / US type |